ABB EasyLine Continuous Gas Analyzer EL3020
Integral air inlet
The integral air inlet (option) is available in two versions. It includes
– Microfilter and flow sensor module
– or microfilter, solenoid valve, pump, coarse filter, capillary tube and flow sensor module.
Electrical interface
Electrical interfaces for outputting measured values and communicating with external systems include Electrical interfaces for communicating with external systems include:
– Integrated Ethernet-10/100BASE-T interface (for service and configuration and configuration) Integrated Ethernet-10/100BASE-T interface dcs-sis.com (for service and configuration purposes) Functional range and ordering
– Analog output module with 2 analog outputs,
– Analog output module with 2 analog outputs, digital input/output module with 4 digital inputs and 4 digital outputs and
Digital input/output module with 4 digital inputs and 4 digital outputs and
– Modbus modules with RS485 and RS232 interfaces.
Housing design
The EL3020 Gas Analyzer is designed in a 19-inch housing with 3 height units and IP20 protection.
Description of Analyzer Performance Characteristics
The performance characteristics of the analyzers are defined according to the international standard IEC 1207-1:1994 “Performance statements for gas analyzers”.
They are based on N2 as a companion gas. When measuring other gas mixtures, compliance with these characteristics can only be ensured if their composition is known.
Measurement principle
Non-dispersive infrared absorption photometers in the λ = 2.5-8 µm wavelength range can measure up to 4 compositions using 1 or 2 beam paths and 1 or 2 receivers per beam path in one gas path or two independent gas paths.
Measure up to 4 components in one gas path or two independent gas paths with 1 or 2 receivers per beam path.
Impact Effects
Flow rate effects
Flow rates in the range of 20-100 liters/hour: within the detection range
Correlated gas effects/cross sensitivity
Analyzer configuration must be aware of sample gas composition.
Selective measures to reduce associated gas effects (optional): Installation of interference filters or filter units, internal electronic cross-sensitivity correction of one sample component by other sample components measured by the gas analyzer.