On October 1, 2017, the Chemicals, Explosives and
Microbiological Hazards Division of Health and
Safety Executive in the United Kingdom issued
Safety Alert Bulletin Number CE MHD 1 2017.
A safety issue was discovered with another
manufacturer’s Gas Chromatograph that uses the
“PX” protection concept in the upper Electronics
Enclosure. This design is based on a purge flow of
air being used to dilute any flammable gasses
present below the flammable limits. This issue was
unfortunately discovered after an instrument
technician was injured as a result of an explosion.
The general concern presented in Safety Alert
Bulletin Number CE MHD 1 2017 is the possibility that
a similar set of circumstances may be replicated
with GC equipment supplied by other manufacturers.