Standard features

Modular design includes

– Modular software – application-based, plug-in software modules

– Manifold module (internal tubing now replaced manifold)

– Analysis section contains stream selection solenoids,

pressure regulation, 32-bit digital detector electronics, and a

single-train chromatograph in a single, replaceable module

– 32-bit digital, low power, controlling electronics. Using

Windows CE® (internal to GC unit)

– Microsoft® Windows® 2000 or XP-based, remote user

interface software

– Lithium battery-backed RAM

– Two remote serial digital communications ports; one local port

– Comprehensive diagnostics and wizards available to users

– Two-level security for user access control

– Audit-quality historical data; date and time stamped

– Operational alarms available with each analysis cycle

– Detectors – constant temperature, glass encapsulated

thermister beads for rugged service and long life. Will not

burn out on loss of carrier

– Valves have no moving metal parts

– Single stream – Manual Calibration

– On demand diagnostics

– Available standard sample conditioning system

– Front panel consisting of LED indicators for: Run, Hold,

Calibrate, Warning, Fault, Warm-up

– USB (host and client) and ethernet ports

– SD memory cards

– Optional Feed-through heater

– AGA3/AGA7 flow calculations

– API 21.2 compliant flow computer functionality