Environmental limits
Electromagnetic compatibility (EMC)
Comply with 2014/30/UE to standards EN 61326-1:2013.
For IEC 61508 SIL certified transmitter to
EN 61326-3-1:2008.
For transmitter with option “YE” to NAMUR NE 021 (2004).
Surge immunity level (with surge protector): 4 kV
(according to IEC 61000-4–5 EN 61000–4–5)
Pressure equipment directive (PED)
Comply with 2014/68/UE to standards
ANSI/ISA 61010-1:2012
Category III Module H for PS ≥ than 20 MPa, 200 bar
Sound Engineering Practice (SEP) for PS < 20 MPa, 200 bar
Humidity
Relative humidity: up to 100 %
Condensing, icing: admissible
Vibration resistance
Accelerations up to 2 g at frequency up to 1000 Hz
(according to IEC 60068–2–6)
Shock resistance
Acceleration: 50 g
Duration: 11 ms
(according to IEC 60068–2–27)